Microlab 300 Semi Automated Chemistry Analyzer: Berikut feature-feature dan spesifikasi alat Microlab 300:
SPECIFICATIONS Microlab 300 Semi Automated Chemistry Analyzer
OPTICAL
• Photometric range: -0.1 to 2.3 Absorbance
• Wavelengths: automatic by 12 positions fi lter wheel
• 7 Standard interference fi lters: 340, 405, 505, 546, 578, 620 and 660 nm
• 5 positions for optional fi lters
• Detector: Photo diode (320-1000 nm)
• Light source: quartz-halogen lamp 12V-20W
TEMPERATURE CONTROL
• Peltier system
• Fixed temperature at 37°C
FLUIDIC SYSTEM
• Internal bellows pump, driven by stepper motor
• Back panel connection for waste
• Aspiration volume programmable
FLOWCELL
• Metal, with quartz windows
• Measuring volume 30 μL
OPERATOR INTERFACE
• Membrane keyboard, for direct function and alpha-numeric entry
• Optional external keyboard
• High contrast graphical LCD display
PRINTER
• Internal matrix printer
• Normal paper
• External printer port available
SIGNAL INTERFACE
• Centronics type parallel port
• RS 232 type serial port
• PS 2 type port for external keyboard
• Real time clock, 24 hours system
LANGUAGES
• English, Spanish, French, German and Portuguese.
• Other languages on request
MEASUREMENT PROCEDURES
• Kinetic, with linearity check
• Kinetic, with linearity check and sample slope blank
• Two point kinetic, with or without reagent blank
• End point, with or without reagent blank
• Bichromatic end point, with or without reagent blank
• End point, with sample blank and with or without reagent blank
• Blanking: automatic zero setting
MULTIPLE TESTING
• Up to 9 replicates
• Means, SD and CV%
MEASURING TIME
• Programmable, 2 to 998 seconds for end point, kinetic and two point assays
DELAY TIME
• Programmable, 0 to 999 seconds
METHOD PARAMETER SETTINGS
• Method name
• Measurement mode
• Wavelength 1 and 2
• Aspiration volume
• Measurement delay
• Measurement time
• Factor
• Concentration standards
• Reagent blank y/n
• Sample blank y/n
• Units for results
• Levels for fl agging
• Curve fi t y/n
• Linearity check
CALIBRATION
• Factor, one-point, two-point and multipoint
• Automatic on 1 standard (linear mode)
• Automatic on up to 10 standards (nonlinear mode)
QUALITY CONTROL
• Two controls per test
• QC survey of last 30 control measurements
• Levey Jenning plot
• High/Low fl ags
STANDARDS AND REGULATIONS
• CE – IVDD
• USA FDA listed
• CB
POWER REQUIREMENTS
• 100-240 Vac nominal, 50/60 Hz.
• Battery back-up to retain data
DIMENSIONS & WEIGHT
• 40 cm (16 in) x 17 cm (7 in) x 36.5 cm (14 in) (W x H x D)
• 8.5 kg (19 lbs).